Successful Lead Abandonment as a Safe Strategy in Complex Pacemaker Lead Entrapment: A Case Report and Review of the Literature

Document Type : Case report

Authors
1 Cardiology department, Mashhad university of medical science, mashhad, iran
2 Vascular & Endovascular Surgery Research Center, Mashhad university of medical sciences, mashhad, iran
3 Cardiology department, mashhad university of medical sciences, mashhad, iran
Abstract
Introduction: Managing dysfunctional pacemaker leads in elderly patients requires careful consideration of the risks associated with extraction versus the potential long-term complications of abandonment.



Case Presentation: An 87-year-old female with a dual-chamber pacemaker implanted two years prior for complete heart block presented with syncopal episodes following heavy lifting. Device interrogation showed intermittent high impedance and elevated threshold in the right ventricular (RV) lead with intermittent capture failure induced by pocket manipulation. Chest radiography revealed no macroscopic lead defect. Extraction of the RV lead using simple traction resulted in dislodgement of the right atrial (RA) lead. Attempts to reposition failed, and subsequent efforts to extract both leads led to their entanglement within the left subclavian vein. Further endovascular extraction attempts were unsuccessful. After discussing surgical risks versus conservative options with the patient and family, the decision was made to abandon the leads. The proximal portions were cut and sutured to the pocket floor, and a new pacemaker system was implanted via the right subclavian vein. At one-year follow-up, the new device functioned normally, and the left subclavian vein remained patent on Doppler ultrasound.



Conclusion: In elderly patients with complex lead entrapment where extraction carries high procedural risk, lead abandonment with careful clinical surveillance offers a safe and effective alternative to invasive removal.
Keywords

1. Chandra MS, Patel R, Singh S. "False inhibition" of demand pacemaker due to leakage of fluid into the pacemaker lead socket. J Thorac Cardiovasc Surg. 1978;75(5):765-8. 2. Bracke F, Rademakers LM, van Veghel D. When pacing or defibrillator leads become redundant: Extract or abandon? Heart Rhythm. 2024 Jul 31:S1547-5271(24)03089-3. doi: 10.1016/j.hrthm.2024.07.113. 3. User259229. Retained cardiac pacemaker leads [Internet]. Radiopaedia.org; 2025 [cited 2026 Feb 23]. Available from: https://radiopaedia.org 4. FDA MAUDE Database. Adverse Event Report: Boston Scientific Corporation Fineline II EZ Sterox; Implantable Lead. Report Number: 2124215-2022-44638. 2022. 5. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, Cha YM, Clancy J, Deharo JC, Ellenbogen KA, Exner D. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart rhythm. 2017 Dec 1;14(12):e503-51. 6. Segreti L, Rinaldi CA, Claridge S, Svendsen JH, Blomstrom-Lundqvist C, Auricchio A, Butter C, Dagres N, Deharo JC, Maggioni AP, Kutarski A. Procedural outcomes associated with transvenous lead extraction in patients with abandoned leads: an ESC-EHRA ELECTRa (European Lead Extraction ConTRolled) Registry Sub-Analysis. EP Europace. 2019 Apr 1;21(4):645-54. 7. Bongiorni MG, Kennergren C, Butter C, Deharo JC, Kutarski A, Rinaldi CA, Romano SL, Maggioni AP, Andarala M, Auricchio A, Kuck KH. The European lead extraction controlled (ELECTRa) study: a European heart rhythm association (EHRA) registry of transvenous lead extraction outcomes. European heart journal. 2017 Oct 21;38(40):2995-3005. 8. Zerbe F, Ponizynski A, Dyckmans J, et al. Functionless retained pacing leads in the cardiovascular system. A complication of pacemaker treatment. Br Heart J. 1985;54(1):76-9. 9. Al-Qudah AAM, Khanfar A, Damra E, Mannan F, Malik N. Coronary sinus dislodgement of pacemaker RA lead tip: a case report. Heart. 2025;111(Suppl 3):A118. 10. Kimura T, Kitamura T, Tokioka S, et al. Cardiac perforation due to a fracture of a recalled Accufix bipolar active fixation pacing lead 29 years after implantation: A case report. J Cardiovasc Electrophysiol. 2021;32(5):1461-3.